Acceso abierto

Research on AC voltage measurement method based on high-speed comparison and Josephson Junction Arrays Voltage Standard

, , ,  y   
21 mar 2025

Cite
Descargar portada

Figure 1.

Schematic diagram of the AC voltage measuring device based on the quantum DC voltage reference
Schematic diagram of the AC voltage measuring device based on the quantum DC voltage reference

Figure 2.

Schematic diagram of the measured sinusoidal voltage signal sampling
Schematic diagram of the measured sinusoidal voltage signal sampling

Figure 3.

Schematic diagram of the simulation circuit
Schematic diagram of the simulation circuit

Figure 4.

Surmograph of the 50 kHz measured sinusoidal voltage signal
Surmograph of the 50 kHz measured sinusoidal voltage signal

Figure 5.

Plot of the reproduced waveform error changes with the measured signal frequency
Plot of the reproduced waveform error changes with the measured signal frequency

Performance comparison among PJVS, JAWS and the method proposed in this paper

Method Amplitude of measured Voltage Frequency Range Accurate measurement of magnitude in optimum condition Measurement System complexity
Presented in this paper 3V 50kHz~500kHz 3×10-5 Low
PJVS 10V DC~1kHz 7×10-7 Medium
JAWS About 1.414V 10Hz~4MHz About 1×10-5 High